Limited official informationFDA Medical Device Enforcement ReportsRecall date: Jul 29, 2026

Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001

Full official product description

Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Processing System.

Verification delayed

The last verified facts remain visible. Last successful verification: Aug 5, 2026, 7:44 AM UTC. A temporary source failure does not mean the recall changed.

Official information is incomplete

Missing: action, consumer scope. The page is public for lookup but excluded from search indexing and advertising.

1Is my item affected?

Compare the complete identifiers. Name similarity alone is not a confirmed match.

Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Processing System.

Identifiers to compare

  • Model M00139509001
Affected variants from the official event record
Product or modelCodes and lotsQuantityRecall number
Boston Scientific TheraSphere 360 Y-90 Management Platform, Model Number M00139509001; Radiological Image Processing System.UDI/DI 00191506068264, All released software versions (V1.1.2, V1.2.1, V1.2.2. V2.0.1, V2.0.2)68 unitsZ-2771-2026

2What is the risk?

The agency's statement is shown without a site-authored severity score.

Software anomaly occurs when a user adds one or more vials to the cart and subsequently returns to the Activity Calculator to modify one or more input. When a user modifies an input and selects "Recalculate", the dose calculation is updated correctly; however, any vials that were previously added to the cart are not automatically updated to reflect the revised calculation.

Agency classification: Class II

3What should I do?

Only instructions associated with this recall event qualify as an action.

Consumer action is not available in the associated official record. The site does not infer or generate a replacement.

Verify on the official source

The linked agency record controls if any information differs.

Official evidence

Official scope

Distribution reported by the agency

US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, IL, IN, LA, MA, MN, NE, NY, OH, OK, OR, SC, SD, TN, TX, UT, WA, WI, WV, WY.

Distribution is context, not a confirmed item match. Compare the model, lot, UPC, NDC, or serial number above.

Material change history

No material affected-item, risk, action, or status change has been recorded since this page entered the tracker.