Limited official informationFDA Medical Device Enforcement ReportsRecall date: Jul 1, 2026

Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 US Retail Sensor/Transmitter 1-pack Model/Catalog Numb...

Full official product description

Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 US Retail Sensor/Transmitter 1-pack Model/Catalog Number: STP-AT-012 Software Version: N/A Product Description: Dexcom G7 US Retail Sensor/Transmitter 1-pack Component: Dexcom G7 Continuous Glucose Monitoring System

Verification delayed

The last verified facts remain visible. Last successful verification: Aug 5, 2026, 7:44 AM UTC. A temporary source failure does not mean the recall changed.

Official information is incomplete

Missing: action, consumer scope. The page is public for lookup but excluded from search indexing and advertising.

1Is my item affected?

Compare the complete identifiers. Name similarity alone is not a confirmed match.

Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 US Retail Sensor/Transmitter 1-pack Model/Catalog Number: STP-AT-012 Software Version: N/A Product Description: Dexcom G7 US Retail Sensor/Transmitter 1-pack Component: Dexcom G7 Continuous Glucose Monitoring System

Identifiers to compare

  • UDI 00386270004109
  • Lot Code
  • Lot Numbers
  • Lot 1725069002
Affected variants from the official event record
Product or modelCodes and lotsQuantityRecall number
Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 US Retail Sensor/Transmitter 1-pack Model/Catalog Number: STP-AT-012 Softw...Lot Code: UDI-DI: 00386270004109 Lot Numbers: 1725204004 and 1725069002N/AZ-2574-2026

2What is the risk?

The agency's statement is shown without a site-authored severity score.

Certain lots of Dexcom G7 sensors originally designated as scrap and intended for destruction were stolen during the destruction process. The sensors were later sold to pharmacies and distributors by Pharmsource, LLC. Lot 1725204004 were designated for destruction because the G7 sensors were not properly sterilized and have an increased risk of skin infection if used. Lot 1725069002 were designated for destruction due to an increased risk of having no sensor readings available or experiencing sensor inaccuracy that includes the potential for missed detection of a hyperglycemic event or incorrect treatment decisions that can result in patient injury, including medical intervention to prevent permanent injury or impairment. Additionally, due to the unknown storage and handling and storage conditions these sensors may have undergone during the distribution, these sensors may present additional risk.

Agency classification: Class II

3What should I do?

Only instructions associated with this recall event qualify as an action.

Consumer action is not available in the associated official record. The site does not infer or generate a replacement.

Verify on the official source

The linked agency record controls if any information differs.

Official evidence

Official scope

Distribution reported by the agency

Dexcom does not have distribution records for this product due to its unauthorized release. However Dexcom is able to identify sensor session starts for the affected lots through the Dexcom G7 CGM App user performance data. Based on App user performance data, Lot 1725204004 has sensor session starts in the United States. Based on App user performance data, Lot 1725069002 has sensor session starts in the United States and the countries of Australia, Austria, Canada, Czech Republic, Finland, France, Germany, Hong Kong, Israel, Italy, Japan, South Korea, Kuwait, Netherlands, Poland, Portugal, Qatar, Singapore, Spain, Switzerland, Turkey, United Arab Emirates,and the United Kingdom.

Distribution is context, not a confirmed item match. Compare the model, lot, UPC, NDC, or serial number above.

Material change history

No material affected-item, risk, action, or status change has been recorded since this page entered the tracker.