Limited official informationFDA Medical Device Enforcement ReportsRecall date: May 6, 2026

Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Catalog Number: CMP-01669 Software...

Full official product description

Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Catalog Number: CMP-01669 Software Version: N/A - not software version specific Component: Yes, the Genius" Review Station is a component of the Genius" Digital Diagnostics System with the Genius" Cervical AI Algorithm cleared under DEN210035.

Verification delayed

The last verified facts remain visible. Last successful verification: Aug 2, 2026, 8:43 AM UTC. A temporary source failure does not mean the recall changed.

Official information is incomplete

Missing: action, consumer scope. The page is public for lookup but excluded from search indexing and advertising.

1Is my item affected?

Compare the complete identifiers. Name similarity alone is not a confirmed match.

Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Catalog Number: CMP-01669 Software Version: N/A - not software version specific Component: Yes, the Genius" Review Station is a component of the Genius" Digital Diagnostics System with the Genius" Cervical AI Algorithm cleared under DEN210035.

Identifiers to compare

  • UDI 15420045514201
  • Model CMP-01669
  • Serial All
Affected variants from the official event record
Product or modelCodes and lotsQuantityRecall number
Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Catalog Number: CMP-01669 Software Version: N/A - not s...Model Number: CMP-01669; UDI-DI: 15420045514201; Serial Number: All;1200 unitsZ-1949-2026

2What is the risk?

The agency's statement is shown without a site-authored severity score.

It was identified that certain devices were being operated outside the validated and FDA-cleared display configuration due to user modification of the manufacturer-installed settings. Modifications included disabling Barco Application Appearance Manager (AAM), altering calibration and compliance parameters, and adjusting additional display-related settings. This field correction is being conducted to restore affected system to the validated configuration and reinforce proper use in accordance with the Operator's Manual.

Agency classification: Class II

3What should I do?

Only instructions associated with this recall event qualify as an action.

Consumer action is not available in the associated official record. The site does not infer or generate a replacement.

Verify on the official source

The linked agency record controls if any information differs.

Official evidence

Official scope

Distribution reported by the agency

Worldwide distribution - US Nationwide and the countries of Asia Pacific, Europe, and the Middle East.

Distribution is context, not a confirmed item match. Compare the model, lot, UPC, NDC, or serial number above.

Material change history

No material affected-item, risk, action, or status change has been recorded since this page entered the tracker.