iLet Bionic Pancreas, REF: BB1001

insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous Glucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, leading to only basal/requested meal doses provided, no correction doses/insulin reductions/suspensions for high/low glucose unless blood glucose entered for sensor life or replacement. Potential health risks to users include severe hyperglyc...

Official details

Category
Medical device recall
Status
Ongoing
Classification
Class II
Company or vendor
Beta Bionics, Inc.
Affected product
US Nationwide distribution in the states of TX, PA, AL, MI, OH, FL, TN, GA.
Risk or reason
insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous Glucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, leading to only basal/requested meal doses provided, no correction doses/insulin reductions/suspensions for high/low glucose unless blood glucose entered for sensor life or replacement. Potential health risks to users include severe hyperglycemia with and without DKA and severe hypoglycemia.
Remedy or action
Software versions 1.4.3, and 1.4.2
Official ID
Z-1996-2026