Solar GI System (Water Perfusion)

Full official product description

Solar GI System (Water Perfusion); Product Name (Catalog Number): Solar GI LGI H2O, 12p, Trolley system (G3-6), Solar GI HRM H2O, 24p, Compact Pole system (G3-7), Solar GI HRM H2O, 24p, Trolley system (G3-8), Solar GI HRM H2O, 36p, Trolley system (G3-12), Solar GI HRAM H2O, 24p, Trolley system (G3-14);

Verification delayed

The last verified facts remain visible. Last successful verification: Aug 5, 2026, 7:44 AM UTC. A temporary source failure does not mean the recall changed.

Official information is incomplete

Missing: action, consumer scope. The page is public for lookup but excluded from search indexing and advertising.

1Is my item affected?

Compare the complete identifiers. Name similarity alone is not a confirmed match.

Solar GI System (Water Perfusion); Product Name (Catalog Number): Solar GI LGI H2O, 12p, Trolley system (G3-6), Solar GI HRM H2O, 24p, Compact Pole system (G3-7), Solar GI HRM H2O, 24p, Trolley system (G3-8), Solar GI HRM H2O, 36p, Trolley system (G3-12), Solar GI HRAM H2O, 24p, Trolley system (G3-14);

Identifiers to compare

  • UDI 08717775955047
  • UDI 08717775955054
  • UDI 08717775955061
  • UDI 08717775955078
  • UDI 08717775955085
  • Serial numbers
Affected variants from the official event record
Product or modelCodes and lotsQuantityRecall number
Solar GI System (Water Perfusion); Product Name (Catalog Number): Solar GI LGI H2O, 12p, Trolley system (G3-6), Solar GI HRM H2O, 24p, Compact Pole system (G...1. Catalog Number: G3-6; UDI-DI: Primary UDI-DI: 08717775955047; Basic UDI-DI: 87177759500G3G8; All serial numbers; 2. Catalog Number: G3-7; UDI-DI: Primary UDI-DI: 08717775955054; Basic UDI-DI: 87177759500G3G8; All serial numbers; 3. Catalog Number: G3-8; UDI-DI: Primary UDI-DI: 08717775955061; Basic UDI-DI: 87177759500G3G8; All serial numbers; 4. Catalog Number: G3-12; UDI-DI: Primary UDI-DI: 08717775955078; Basic UDI-DI: 87177759500G3G8; All serial numbers; 5. Catalog Number: G3-14; UDI-DI: Primary UDI-DI: 08717775955085; Basic UDI-DI: 87177759500G3G8; All serial numbers;372 unitsZ-2669-2026

2What is the risk?

The agency's statement is shown without a site-authored severity score.

Under extended clinical use conditions (e.g., long duration studies), moisture generated within the pneumatic circuit may accumulate inside the system's internal tubing or at the hydrophobic filter and waterproof pressure sensors. Accumulated moisture may affect airflow or pressure measurement performance if not properly removed.

Agency classification: Class II

3What should I do?

Only instructions associated with this recall event qualify as an action.

Consumer action is not available in the associated official record. The site does not infer or generate a replacement.

Verify on the official source

The linked agency record controls if any information differs.

Official evidence

Official scope

Distribution reported by the agency

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Canada, China, Costa Rica, Croatia, Cyprus, Czech Republic. Denmark, Egypt, France, Germany, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Libya, Lithuania, Malaysia, Namibia, Nepal, Netherlands. Oman, Osterreich, Pakistan, Poland, Portugal, Puerto Rico, Qatar, Russia, Russian Federation, Saudi Arabia, Serbia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Tunisia, Turkey, UAE, United Kingdom, and Vietnam.

Distribution is context, not a confirmed item match. Compare the model, lot, UPC, NDC, or serial number above.

Material change history

No material affected-item, risk, action, or status change has been recorded since this page entered the tracker.